Risø High Dose Reference Laboratory - HDRL

Risø HDRL services the radiation processing industry in Denmark and abroad, and in particular the medical device industry that sterilizes products using radiation.

 
Accreditation

Risø HDRL is accredited by DANAK – Danish Accreditation in accordance with EN ISO 17025:2005 for calibrations and dose measurements in the high-dose range 0.2 – 100 kGy.

Accreditation document,  link to PDF file.

Accreditation information:
http://published.danak.dk/register.asp?akk=266&lang=e&nohead=y

Specific measurement areas and measurement uncertainties:
http://published.danak.dk/showdata.asp?schema=kalibreringsdata&lang=e&akk=266


Risø HDRL services:

Calibrations are carried in accordance with the recommendations given in the report CIRM 29 (Guidelines for the calibration of dosimeters for use in radiation processing). Link to PDF file.

1. Irradiation of dosimeters and samples to specified doses. Irradiations are carried out at the Risø HDRL cobalt-60 gamma-cells, at a 10-MeV electron accelerator or at the Risø HDRL 100 keV electron accelerator.

2. Calibration of dosimeters by irradiation of the dosimeters at the user’s facility together with HDRL reference transfer dosimeters.
Risø HDRL reference transfer dosimeters are
dichromate solution in ampoules
- alanine pellets and films

3. Issue of calorimeters for routine or reference dose measurement at the user’s electron beam facility. Risø HDRL supplies the software CalDose for dose measurement with calorimeters.

4. Dosimetric validation of irradiation facilities for Installation and Operational Qualification (IQ – OQ) in accordance with the requirements in the international standard for radiation sterilization EN ISO 11137-1:2006.

5. Measurement of dose distribution in products for performance qualification (PQ) in accordance with the requirements in the international standard for radiation sterilization EN ISO 11137-1:2006.

6. Risø B3 radiochromic dosimeter film for dose mapping.

7. RisøScan – software for dose measurement with radiochromic dosimeter film and an office scanner. [Link to Risø HDRL website]

8. Other dosimetry equipment
- aluminium wedge for energy determination at electron accelerators,
- phantoms for irradiation of dosimeters at gamma and electron accelerator facilities.

9. Courses on 
Validation and Process Control for Low Energy Electron Beam Irradiation 

Validation and Process Control for Electron Beam Sterilization